Flow rate is one of the most visible hydrogen-machine specifications—and one of the easiest to misread. A value in ml/min describes gas volume per unit time. It does not by itself establish gas purity, inhaled concentration, treatment dose, clinical effect or suitability for a particular user.
What ml/min does and does not tell you
A machine rated at 300 ml/min nominally generates 0.3 litres of the specified gas stream per minute at the stated operating point. Over one hour, that corresponds to 18 litres of generated gas if output remains constant. This is a generator-volume calculation—not the amount retained by a person and not the percentage in every breath.
Actual inhaled concentration is influenced by the gas composition, nasal-cannula delivery, room air entrainment, breathing pattern, leaks and ventilation. Product pages should therefore keep generator output separate from exposure concentration.
Model-level comparison
| Model | Published output | Outlet/configuration note | What to reconfirm |
|---|---|---|---|
| H300 | 100–300 ml/min H2 | Single outlet | Adjustment steps, tolerance and purity test |
| H300AD | 100–300 ml/min H2 | Single/dual wording must be resolved by configuration sheet | Outlet count and total-versus-per-outlet flow |
| H600 | 600 ml/min H2, listed ±5% | Single or dual option | How flow is divided with two outlets |
| HO1000 | 1000 ml/min total H2/O2, nominal 2:1 | Mixed-gas architecture | Component flows and ratio tolerance |
| HO1000L | 1000 ml/min total H2/O2, nominal 2:1 | High-flow mixed-gas system | Current revision, outlet and gas analysis |
How should a factory test flow?
- Stabilise the unit: use the specified water and warm-up period.
- Define conditions: record ambient pressure and temperature or state the reference correction.
- Use a suitable meter: the instrument must match the gas, expected range and back-pressure.
- Test every setting: record minimum, intermediate and maximum modes.
- Test each outlet: for a dual-outlet version, measure individual branches and total output.
- Repeat across samples: a single golden sample cannot define production consistency.
Flow, purity and stability are separate specifications
Increasing current can increase electrolysis output, but finished-unit performance also depends on water supply, stack temperature, separators, pressure control and power electronics. A professional QC file should show flow stability over time, purity or composition, outlet moisture, abnormal-condition response and post-ageing performance.
How a distributor should choose a model
- H300: compact adjustable pure-hydrogen positioning where a single outlet is required.
- H300AD: request the exact outlet configuration before quotation or advertising.
- H600: higher pure-hydrogen generator flow and optional outlet configurations.
- HO1000/HO1000L: hydrogen–oxygen architecture; compare total and component flows, not the headline number alone.
Market classification, instructions, electrical requirements and permitted claims also influence the choice. Flow is only one line of the sourcing specification.
Frequently asked questions
Does dual outlet mean double the flow?
Not necessarily. Many systems divide a fixed total output. Obtain total and per-outlet measurements for the exact configuration.
Does a higher flow guarantee a better result?
No. A higher generator flow is an engineering characteristic, not a health-outcome guarantee.
Should a catalogue say “1000 ml/min hydrogen” for a 2:1 H2/O2 machine?
No, unless 1000 ml/min is specifically the measured hydrogen component. If it is total mixed-gas flow, label it as such and state the ratio.
Reference
Explore the hydrogen inhalation technology series
Get a flow-matched model quotation
Include the gas type, required flow, outlet count, destination country, quantity, voltage and plug in the Youborui distributor RFQ form.
Hydrogen Inhalation Engineering Library
Continue with these model-linked verification, safety and sourcing guides: