Quick answer: Waveform names such as sine, square or sawtooth are not enough. A buyer should request the measured magnetic-field trace at the user position, together with amplitude, rise time, time base, probe bandwidth and test conditions.
Slew rate describes how quickly a measured quantity changes with time and is commonly expressed for a magnetic field in tesla per second. It is related to the slope of the measured field trace, but it should not be marketed as a universal effectiveness score.
Waveform label versus measured magnetic field
The controller may generate an electrical command that looks square, but coil inductance, resistance, drive electronics, protection components and load conditions shape the final magnetic field. The useful question is therefore not only ‘What waveform is selected?’ but ‘What field was measured where the user’s foot is positioned?’
- Waveform is the time-domain shape of the signal; the real output may differ from the controller command.
- Rise time records how long the field takes to move between defined percentages of its amplitude.
- Slew rate is a change in field divided by a change in time; the calculation interval must be disclosed.
- Probe bandwidth and sample rate must be high enough to resolve the pulse without hiding fast edges.
- Overshoot, ringing and baseline drift should be visible rather than cropped from the trace.
Specification sheet: what should be disclosed?
| Trace identity | Model, serial, PCB/firmware, level and operating mode |
|---|---|
| Amplitude | Peak or RMS field with unit and uncertainty |
| Timing | Time base, pulse width, repetition rate and duty cycle |
| Rise/fall | Threshold definition and calculated T/s where used |
| Probe | Type, bandwidth, orientation, calibration and distance |
| Stability | Start, mid-session and end-of-session comparison |
Interpretation boundary: A higher calculated slew rate is not automatically better and does not prove a health benefit. It is one engineering descriptor that must be evaluated with field strength, frequency, coil design, exposure and intended use.
How to request evidence that another engineer can reproduce
A screenshot without axes or test conditions is weak evidence. Ask the factory for raw or high-resolution records and a short method statement so your own laboratory can repeat the test.
- Define a measurement coordinate above the left and right active zones.
- Record the magnetic field, not only the voltage driving the coil.
- Include horizontal and vertical scales, probe orientation and instrument settings.
- Capture every user-selectable level and any burst or programme transitions.
- Repeat after thermal stabilisation and across several production samples.
- Set acceptance tolerances in the signed OEM specification and final QC plan.
How the question applies to the Youborui P91-P93 range
P91-family and P92/P93 model pages describe controls and current supplier parameters. Those values should be connected to a model-specific measured output package. P91A and P91C should not inherit a P91 trace by assumption, and P92/P93 should remain separately identified even where rated input power is similar.
RFQ questions for importers and distributors
- Is the published waveform measured magnetically or copied from a controller schematic?
- What percentage thresholds define rise time and slew-rate calculation?
- Can the supplier provide traces for all levels and both foot positions?
- Does output remain within tolerance during the longest intended session?
- Which production test detects coil placement or winding errors?
Search and AI visibility: publish facts, not unexplained numbers
Use headings that answer ‘what is PEMF slew rate’ and ‘how is PEMF waveform measured,’ then connect the explanation to verifiable supplier questions. This earns long-tail traffic without claiming that one waveform is medically superior.
A strong product page defines abbreviations, states units, separates measured specifications from general education and links every model claim to the relevant test condition. This structure helps buyers, search engines and AI systems understand the product without inventing missing details.
Frequently asked questions
Is slew rate the same as frequency?
No. Frequency describes repetition per second; slew rate describes how quickly the measured field changes.
Does a square-wave setting guarantee a square magnetic field?
No. The coil and drive circuit shape the actual magnetic output, which should be measured.
What unit is used for magnetic-field slew rate?
Tesla per second (T/s) is commonly used, with the calculation method and interval stated.
Sources and further reading
The independent buyer-guide sources below informed the comparison questions. Official standards and regulator guidance define measurement, safety, usability or advertising boundaries; they do not certify or endorse Youborui products.
- The PEMF Podcast: PEMF slew-rate guide
- BIPM: SI Brochure
- ICNIRP: low-frequency electromagnetic-field guidelines
- IEC 62311:2019 human-exposure assessment
- FDA: Electromagnetic Compatibility of Medical Devices
- IEC 60601-1-2 electromagnetic disturbances
Continue your technical evaluation
Request a model-specific PEMF OEM/ODM package
Send the model, destination country, voltage, plug, quantity, branding, sales channel and required documents through the Global Distributor & OEM RFQ Center. You can also review our quality-control and export guide or contact the Youborui export team.
Editorial note: Review the Youborui editorial and research policy. Final specifications and compliance documents must be confirmed against the ordered production revision.